Tapped Density Tester SVM II

Request a quote
Create PDF

Language:

The SVM II is our latest generation of tapped density tester. With unprecedented flexbility and intelligent features, it makes tapped density testing as easy and reliable as never before.

The 7“ high resolution touch display with a modern user interface focuses the user on the most important task the new SVM II has to fullfill - 100% USP/EP compliant tapped density testing. With the automatic calculation features of the TestAssist, the user is guided step-by-step through the testing process as quickly as possible. It also automatically calculates the Hausner Ratio and compressibility index and prints the reports after every test.

In addition to its ease of use, the SVM II is the most flexible tapped density tester we ever built. The newly designed SwitchPlate system enables 5 different versions of the SVM II, with USP 1, USP 2 or USP 3 in either one or two test station configurations or a combination with both methods. And the best part - they can also be upgraded at a later time point!

SVM II also supports a variety of method and norms: From USP 1 and USP 2 to ASTM B527-15 and DIN EN ISO 787-11:1995 - with the SVM II, users are on the safe side of valid and conform tapped density testing.

Tapped Density Tester SVM II 5-in-1 Tapped Density Tester

One Device, 5 Variations

The SVM II offers the same flexibility as its 5 predecessor models. It can test for USP method 1, 2 and 3 and can be equipped with either one or two test stations of the same method (USP 1, 2) or a combination of both. USP 3 comes with separate attachment parts and can be installed on station one or two and used with any SwitchPlate. Optional changing from one variant to another can be done by the user within minutes. This makes the SVM II the most flexible tapped density tester we ever built.

Intelligent SwitchPlate System

With the innovative SwitchPlate system, the SVM II is as flexible as ever before. Available SwitchPlates:

  • USP 1 with 1 station
  • USP 1 with 2 stations
  • USP 2 with 1 station
  • USP 2 with 2 stations
  • 2 station combination with USP 1 and USP 2

Modern Touch Interface With TestAssist

The SVM II is controlled by a high resolution touch display with our modern ERWEKA user interface. With its innovative TestAssist, the user is guided through the tapped density testing process from start to finish. The TestAssist automatically computes Tapped Density, Hausner Ratio and Compressibility Index - all the user has to do is read the values from the cylinder and enter them when asked. This reduces errors and delivers results as fast as possible. SVM II is also equipped with the DirectHelp feature - always giving help and guidance at the touch of a button.

1, 2, Test!

The touch interface of the SVM II focuses the user on the most important task to be done with the SVM II: tapped density testing. The central part is the TestAssist, an intelligent assistant with real time feasibility checks that provides support and guarantees the conformity of the entered parameters according to USP/EP/JP. Tapped density testing thus becomes simple and safe. But we didn‘t stop at TestAssist - all other functions, such as the control panel, the qualification menu or the service menu, are also designed to connect user and hardware as quickly as possible.

USP methods 1, 2, 3, ASTM B527-15 and DIN ISO 787-11:1995

Intuitive touch interface

One or two test stations

Tapped Density Tester SVM II SwitchPlates

ERWEKA offers five different SwitchPlates including a license key to purchase with the new SVM II. All SwitchPlates can be interchanged, if requirements change, it is possible to upgrade every SVM II with a new SwitchPlate. The SwitchPlate system offers maximum flexibility and makes every SVM II absolutely futureproof!

Installing a SVM II SwitchPlate is a two step process - first the user changes the plate on the SVM II by fixing it with 6 small screws. The next step is done in the configuration panel - the activation code needs to be entered, which is part of the purchase of a SwitchPlate. This activates the specific SwitchPlate in the TestAssist and report engine of the SMV II.

The following SwitchPlates are available:

This plate has 1 test station and is able to test according to USP method 1

This plate has 2 test stations and is able to test according to USP method 1

This plate has 1 test station and is able to test according to USP method 2

This plate has 2 test stations and is able to test according to USP method 2

This plate has 2 test stations and is able to test either according to USP method 1 or USP method 2

Tapped Density Tester SVM II Accessories

Noise Reduction Box

The ERWEKA noise reduction box is the perfect addition to our tapped density tester. The noise box is made of sturdy polymer housing and is lined with a 30 mm thick, white EPE damping material. It reduces the noise of a tapped volume tester by up to 25 dB (A).

Tapped Density Tester SVM II Technical Data

Weight17 kg
Dimensions (H x W x D)196.5 x 300 x 326 mm
Voltage115/230 V; 50/60 Hz
Fuses1 A
InterfacesUSB A/B, Ethernet, RS232
Protection classI/EN 61140
Protection typeIP 21/IEC 529
Control7" high resolution touch display
Test stations1 or 2 test stations using 5 different SwitchPlates
Supported methods- USP 1: 300±15 taps/min; tap height 14±2mm
- USP 2: 250±15 taps/min; tap height 3±0.2mm
- USP 3 as described in USP <616>
- EP 2.9.34 Bulk density and tapped density of powders
- USP <1174> Powder Flow
- ASTM B527-15 Tap density of metal powders and compounds
- DIN EN ISO 787-11:1995 General test methods, pigments and fillers
Data storage10 reports (temporarily)
ExportNetwork export via external software (Export manager) as XML / PDF
Ambient temprature during operation+10 °C up to +30 °C
Storage and transport temperature+5 °C up to +40 °C
Relative humidity25-80 % non condensing