Search

DT 950 & DT 9510 Series Digital Dissolution Offline System

Request a quote
Create PDF

Language:

Digital Dissolution Offline System DT 950 & DT 9510 Series Key Features

100% USP / EP / JP Compliant

USP Methods 1, 2, 5 and 6

Intuitive Touch Interface

Upgradeable at Any Time

TestAssist for Digital Dissolution Offline Testing

The intelligent TestAssist has been updated for digital offline testing:

  • Easy sample timing configuration
  • Advanced offline system settings
  • Updated test interface for more information and transparency in the test procedure

User Management

The updated user management enables flexible access control for the digital dissolution system:

  • Easy-to-use user rights management
  • Password complexity guidelines for safe access control

Stand-Alone Control of Connected Hardware

The digital embedded PC technology in the digital dissolution tester DT 950 and DT 9510 take full control of the connected system hardware: The pump, either PVP or IPC, the automated sampling station ASS-9/-18 and the sample collector. An external computer is not required, saving space, energy and maintenance, and thanks to the digital technology, without loosing comfortable features and USP/EP/JP compliance. Of course, the offline system includes all the other software features of the DT 950 platform, including method memory, sampling station with temperature sensors in each vessel, USB update and FTP data export.

Technical Data

Weight42 kg
Dimensions (H x W x D)850 x 650 x 650 mm
Voltage115/230 V; 50/60 Hz
Speed20-250 U/min
Vessel volume400 ml / 1000 ml / 2000 ml
Interfaces1x RS-232, 2x USB, 2x Ethernet/RJ45
Test stations (DT 956)6 in 2 rows
Test stations (DT 957)7 in 2 rows
Test stations (DT 958)8 in 2 rows
USP methodsUSP 1 / USP 2 / USP 5 / USP 6
Fuses2 A
Protection classI/EN 61140
Protection typeIP 21/IEC 529
OperationTouchscreen 7", 800x480 Pixel
Sampling positionsHigh-head / Low-head / Cleaning mode
Ambient temperature during operation+10 °C to +30 °C (ambient temperature min. -5 °C below set temperature)
Storage & Transport temp.+5 °C to +40 °C
Relative humidity25-80 % non condensing

Technical Data

Weight110 kg
Dimensions (H x W x D)850 x 1062 x 650 mm
Voltage115/230 V; 50/60 Hz
Speed20-250 U/min
Vessel volume400 ml / 1000 ml / 2000 ml
Interfaces1x RS-232, 2x USB, 2x Ethernet/RJ45
Test stations (DT 9512)12 in 2 rows
Test stations (DT 9513)13 in 2 rows
Test stations (DT 9514)14 in 2 rows
USP methodsUSP 1 / USP 2 / USP 5 / USP 6
Fuses2 A
Protection classI/EN 61140
Protection typeIP 21/IEC 529
OperationTouchscreen 7", 800x480 Pixel
Sampling positionsHigh-head / Low-head / Cleaning mode
Ambient temperature during operation+10 °C to +30 °C (ambient temperature min. -5 °C below set temperature)
Storage & Transport temp.+5 °C to +40 °C
Relative humidity25-80 % non condensing

Technical Data

Weight25 kg
Dimensions (H x W x D)585 x 700 x 515 mm
Channels6,7 or 8
Max. capacity (not for 25 ml test tubes)26 sample intervals
ValveIntegrated 3 way valves
FRL racks26 x 8 / 1.5 ml HPLC vials, 26 x 8 / 4.0 ml HPLC vials, 26 x 8 / 10 ml test tubes, 18 x 8 / 25 ml test tubes
Voltage +- 10%115 - 250 VAC; 50 / 60 Hz

Technical Data

Weight28 kg (for PVP 1220/1420)
Dimensions (H x W x D)420 x 275 x 575 mm (for PVP 1220/1420)
Voltage115 V or 230 V, 50/60 Hz
Pump typePVP 1220/1420 (for DT 9510)
Channels12 or 14 (for PVP 1220/1420)
Valves/
Accuracy+/- 0.5 ml
System compatibilityDT Online System, DT Offline System, DT On-/Offline System
BenefitsFiltration down to 0.22 μm with a flat membrane filtration. Particularly suitable for fully automatic dissolution systems.

Technical Data

Dimensions (H x W x D)125 x 145 x 220 mm
InterfacesRS 232
Channels8 or 16
Accuracy25 ml +/− 5%
Media replacementStandard
Double filtration (optional)Only when first filtration with poroplast filters. No media replacement possible when double filtration.
Required type of sample collectorFRL 654 / 754 / 854
System compatibilityDT Offline / DT Online / DT On-/Offline
AdvantagesBasic pump possible with DT 950/9510, needs regular replacement of tubing